FDA Cybersecurity Rules for Medical Devices

Sabyrix Team November 23, 2026

If your product connects to the internet and runs software that could be exposed to a cybersecurity threat, the FDA may classify it as a "cyber device," and that label comes with legal obligations that have nothing to do with whether the FDA considers your software safe or effective. Section 524B of the Federal Food, Drug, and Cosmetic Act requires a vulnerability disclosure plan, ongoing patch processes, and a software bill of materials before a submission is even accepted for review. These requirements took effect in 2023, and FDA's guidance on how to satisfy them has already been updated twice, most recently in February 2026. This article explains what a cyber device is, what Section 524B actually requires, how it changes the premarket submission process, and what it means for teams building connected diagnostic tools, clinical decision support software, or AI-enabled monitoring products.

This is a different question from whether your software is regulated as a medical device at all. If you have not yet worked through that classification, start with our guide on how FDA decides whether software counts as a medical device. This article picks up from there: once a product is a regulated device and also meets the "cyber device" test, Section 524B adds a separate, specific set of security obligations.

What Makes Software a "Cyber Device"

Section 524B, codified at 21 U.S.C. 360n-2, defines a cyber device as one that meets all three of the following conditions:

  • It includes software that the sponsor validated, installed, or authorized as part of the device.
  • It has the ability to connect to the internet.
  • It contains technological characteristics that could be vulnerable to cybersecurity threats.

In practice, that third condition is rarely the limiting factor. Almost any networked software has some attack surface, whether that is an API endpoint, a cloud sync process, or a firmware update channel. The practical test most regulatory teams apply is simpler: if the device includes software and talks to the internet in any way, assume it is a cyber device until a qualified regulatory reviewer says otherwise.

This catches more products than it used to. A connected glucose monitor, a cloud-synced ECG patch, an AI model that pulls imaging data from a hospital network, and a mobile app that controls an infusion pump can all qualify. The classification is about the device's connectivity and software content, not its risk class. A Class II device and a Class III device can both be cyber devices if they meet the three-part test.

What Section 524B Actually Requires

Section 524B applies to premarket submissions under FD&C Act sections 510(k), 513, 515(c), 515(f), and 520(m), which covers 510(k) clearances, PMA applications, De Novo requests, and HDE submissions, along with certain supplements. For a covered submission, the sponsor must address three things.

A plan to monitor, identify, and address vulnerabilities

The sponsor must submit a plan describing how it will monitor for cybersecurity vulnerabilities and exploits after the device reaches the market, identify and address them within a reasonable time, and run a coordinated vulnerability disclosure process. In practice, this means having a defined intake channel for security researchers, a triage process, and a documented timeline for remediation, not just a promise to "monitor security."

Processes that give reasonable assurance of cybersecurity

The sponsor must design, develop, and maintain processes and procedures that provide reasonable assurance that the device and the systems it connects to are cybersecure. This extends beyond the device itself to the cloud backend, companion apps, and any third-party services the device depends on. FDA also expects a defined patch cadence: a regular cycle for routine, non-critical vulnerabilities, and an expedited, out-of-cycle process for critical vulnerabilities that could create uncontrolled risk.

A software bill of materials

The sponsor must provide FDA with a software bill of materials, commonly called an SBOM, covering the commercial, open-source, and off-the-shelf software components used in the device. This is the requirement that catches the most teams off guard, because it assumes a level of dependency tracking that many software organizations, inside and outside healthcare, have never formally built.

What an SBOM Needs to Actually Cover

An SBOM is not a marketing document or a high-level architecture diagram. FDA's current guidance expects a structured, ideally machine-readable inventory that identifies each software component in the device, along with enough detail to assess its exposure to known vulnerabilities. For most product teams, that means:

  • Every open-source library and package, with version numbers, not just "latest."
  • Commercial and off-the-shelf software components, including embedded operating systems and third-party SDKs.
  • A process for regenerating the SBOM automatically as dependencies change, rather than compiling it by hand before each submission.
  • A way to cross-reference components against vulnerability databases such as the National Vulnerability Database, so a newly disclosed CVE in a dependency can be triaged quickly.

Teams that already use dependency scanning tools in their CI/CD pipeline have most of the raw data already. The gap is usually turning that data into a submission-ready artifact and keeping it current after launch, since the SBOM obligation does not end once the device clears review.

How This Changes the Premarket Submission Process

Section 524B took effect on March 29, 2023, for covered submissions filed on or after that date. FDA's refuse-to-accept policy for incomplete cybersecurity information took effect later that year, on October 1, 2023. From that date forward, 510(k) submissions filed through the eSTAR template can be placed on technical screening hold if the cybersecurity section is incomplete, which stops the review clock before substantive review even begins.

This has a direct practical consequence for project timelines. Cybersecurity documentation is not something to assemble in the weeks before a submission deadline. It needs to be built alongside the product, with the vulnerability monitoring plan, patch process, and SBOM maturing as the architecture does. Teams that treat it as a late-stage compliance checklist item consistently run into submission delays.

FDA's current guidance document, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, was last updated on February 3, 2026, superseding a June 2025 version that itself replaced the original 2023 guidance. The pace of revision is itself worth noting: this is an area FDA continues to actively refine, and guidance that was accurate a year ago may already be out of date.

Cybersecurity Obligations Don't End at Clearance

Section 524B's requirements are not a one-time premarket checkbox. The vulnerability monitoring plan, patch cadence, and SBOM maintenance are ongoing obligations that continue for as long as the device is marketed. FDA's separate postmarket guidance expects manufacturers to use a structured, risk-based approach to triage newly discovered vulnerabilities, decide whether a fix needs a new submission, and communicate with providers and patients when a risk is significant.

Most cybersecurity-only changes, such as patching a dependency or rotating a cryptographic key, do not require a new premarket submission. FDA expects manufacturers to make a documented, risk-based judgment call on that question for each change, not assume every patch is submission-free by default.

What This Means for Teams Building Connected Health Software

If you are scoping a product that might qualify as a cyber device, the Section 524B requirements should shape the architecture from the start, not get bolted on before a submission. A few practical implications:

  • Build dependency tracking into the pipeline early. Generating an SBOM after the fact, across a mature codebase with years of accumulated dependencies, is far more expensive than automating it from the first sprint.
  • Design the backend, not just the device, for cybersecurity. FDA's "related systems" language means the cloud services, APIs, and companion apps around a device are in scope, not just the embedded firmware.
  • Put a coordinated vulnerability disclosure process in place before launch. This typically means a published security contact, a defined intake and triage workflow, and a committed response timeline, which also tends to improve how quickly real issues get fixed regardless of regulatory pressure.
  • Separate the SaMD classification question from the cyber device question. A product can be a regulated medical device without connecting to the internet, and it can connect to the internet without being a regulated device. Only the overlap of both triggers Section 524B.
  • Loop in regulatory counsel before finalizing scope. The statutory test is specific, but applying it to a real product, especially one with AI components or third-party integrations, usually benefits from a reviewer who has done it before.

This is also where the line between healthcare software generally and regulated medical device software matters most. Many patient-facing applications, from scheduling tools to patient portals, handle protected health information without ever becoming a cyber device under Section 524B. HIPAA's security requirements still apply, and the engineering discipline around audit logging and risk analysis overlaps with what FDA expects, but they are legally distinct obligations that need to be tracked separately.

Frequently Asked Questions

Does Section 524B apply to software that FDA does not regulate as a medical device?

No. Section 524B only applies to a covered premarket submission for a device, meaning the product has to clear the medical device threshold first. A patient portal, scheduling app, or general wellness tool that is not a regulated device does not trigger these requirements, even if it connects to the internet.

How is a cyber device different from Software as a Medical Device (SaMD)?

SaMD classification asks whether software's intended use makes it a regulated medical device at all. Being a cyber device is a separate, narrower question that only applies once a product is already a regulated device: does it include software, connect to the internet, and carry characteristics that could be vulnerable to a cybersecurity threat. A device can be SaMD without being a cyber device, for example if it never connects to a network.

Do routine security patches require a new FDA submission?

Usually not. FDA's guidance recognizes that most changes made solely to address cybersecurity, such as patching a known vulnerability, typically do not require a new premarket submission. Manufacturers still need a documented, risk-based process to make that determination for each change rather than assuming it by default.

What should a startup building a connected diagnostic device do first?

Confirm the SaMD classification, confirm whether the product meets the three-part cyber device test, and then build the vulnerability monitoring plan, patch process, and SBOM generation into the engineering plan before architecture decisions are locked in. Retrofitting these processes onto a near-finished product is significantly more expensive than designing for them from the start.

Is this article legal or regulatory advice?

No. This is a technical and software engineering explanation of a regulatory framework, not legal advice. FDA regulatory classification and submission strategy depend on the specific device, its intended use, and current guidance, and should be confirmed with qualified regulatory counsel before you rely on it for a real submission.

Section 524B adds real engineering work, not just paperwork, to any connected health product that crosses into device territory. Teams that plan for vulnerability monitoring, patch processes, and SBOM generation from the start of a build avoid the submission delays that come from treating cybersecurity as an afterthought. If you are scoping a connected diagnostic tool, an AI-enabled clinical product, or any software that might cross the cyber device threshold, our healthcare AI development team can help you work through the architecture and compliance implications together, or you can book a strategy call to talk through your specific product.